Oscillometric blood pressure devices

Primary objective: To measure the repeatability and pressure pulse envelope of simulators used for testing oscillometric non-invasive blood pressure (NIBP) devices; to study the effect of different envelopes on NIBP devices, and to measure the difference between NIBP devices due to different oscillometric algorithms.

Methods: Three different models of NIBP simulator and 18 different patient monitors with NIBP function were studied. We developed a pressure measurement system (accuracy 0.048?mmHg) to measure the repeatability of simulators. The effect of changing the envelope was measured by using three simulators with one NIBP device. Differences between 18 NIBP devices were measured using one simulator at seven blood pressure settings.

Main outcomes and results: Simulators generate repeatable pressure pulse envelopes (<0.2mmHg) but the magnitude and shape depends on the model of simulator. Oscillometric NIBP devices are highly repeatable (<2mmHg) when presented with a repeatable pressure pulse envelope, but different devices employ different algorithms and give different results. For a simulated standard blood pressure setting of 120/80?mmHg, estimates of systolic pressure ranged from 112.6 to 126.6mmHg (sd of 3.0mmHg), and diastolic pressure ranged from 74.8 to 86.9?mmHg (sd of 3.5mmHg).

Conclusions: Simulators and NIBP devices are sufficiently repeatable for clinical use, but further systematic clinical studies are required to better characterize the pressure pulse envelope for different patient groups.

Publication

Sims AJ, Reay CA, Bousfield DR, Menes JA, Murray A. Oscillometric blood pressure devices and simulators: measurements of repeatability and differences between models. Journal of Medical Engineering & Technology 2005; 29: 112-118.

Contact

Dr. Andrew J. Sims, Freeman Unit, Newcastle upon Tyne. Tel: +44 (0)191-223-1370.

Top of page

W3C XHTML 1.0 logo
W3C CSS logo
Triple-A conformance icon, W3C-WAI Web Content Accessibility Guidelines 1.0
Internet Content Rating Association logo

Regional Medical Physics Department, Newcastle General Hospital, Newcastle upon Tyne, NE4 6BE, UK.
Tel: +44 (0)191-233-6161 Fax: +44 (0)191-226-0970 E-mail: webmaster@rmpd.org.uk Web: http://www.rmpd.org.uk
By using this site you agree to our Terms and Conditions of Use. Please read our legal page.
Regional Medical Physics Department, a Clinical Directorate of the Newcastle upon Tyne Hospitals NHS Foundation Trust.
Copyright ©1999-2008 All Rights Reserved. Revised 28 April, 2008.

This Web Site

Accessibility | Feedback | Help | Links | Site Map | What's New |

Our Organisation

FAQs | Find Us | News & Events | Services | Work For Us |

Our Research

Abstracts | Innovation | Publications | What We Do |

Quick Site Search

Advanced search