Automated non-invasive blood pressure measuring devices based on the oscillometric technique are used widely for self-measurement and are often used in clinics in place of the manual, auscultatory method. Oscillometry was originally developed for monitoring purposes and there are questions over its suitability for making diagnostic measurements.
This study measured the differences between automated devices, in the absence of physiological variability. We studied 19 low-cost, automated, non-invasive blood pressure devices, using a repeatable artificial arm simulator, and measured the within-device repeatability and between-device differences. We found that the devices were repeatable (mean within-device difference 1 mmHg), but between-device differences were 4.4 mmHg (systolic pressure) and 3.6 mmHg (diastolic pressure), for normal and high-normal blood pressures. Individual devices are sufficiently repeatable for clinical trend use, but differences between devices are sufficiently large that they may be misinterpreted as clinically significant.
Sims AJ, Reay CA, Bousfield DR, Menes JA, Murray A. Low-cost oscillometric non-invasive blood pressure monitors: device repeatability and device differences. Physiol Meas 2005; 26: 441-445.
Dr. Andrew J. Sims, Freeman Unit, Newcastle upon Tyne. Tel: +44 (0)191-223-1370.
Regional Medical Physics Department, Newcastle General Hospital, Newcastle upon Tyne, NE4 6BE, UK.
Tel: +44 (0)191-233-6161 Fax: +44 (0)191-226-0970 E-mail: webmaster@rmpd.org.uk Web: http://www.rmpd.org.uk
By using this site you agree to our Terms and Conditions of Use. Please read our legal page.
Regional Medical Physics Department, a Clinical Directorate of the Newcastle upon Tyne Hospitals NHS Foundation Trust.
Copyright ©1999-2008 All Rights Reserved. Revised 28 April, 2008.