Comparison of magnetocardiography and electrocardiography

Objective: Automated techniques were developed for the measurement of cardiac repolarisation using magnetocardiography.

Methods: This was achieved by collaboration with the Physikalisch-Technische Bundesanstalt (PTB), Berlin, Germany and the Grönemeyer Institute of Microtherapy, Bochum, Germany, to obtain recordings of magnetocardiograms (MCGs) in cardiac patients and healthy subjects. Manual and automated ventricular repolarisation measurements from MCGs were evaluated to determine the clinical relevance of these measurements compared with electrocardiograms (ECGs).

Results: Results showed that MCG and ECG T-wave shapes differed and that manual repolarisation measurement was significantly influenced by T-wave amplitude. Automatic measurements of repolarisation in both MCGs and ECGs differed between techniques. The effects of filtering on the waveforms showed that filtering in some MCG research systems could significantly influence the results, with 20 ms differences common. In addition, MCGs were better able to identify differences in the distribution of cardiac magnetic field strength during repolarisation and depolarisation between normal subjects and cardiac patients. Differences were also determined in ventricular repolarisation between MCGs and ECGs, which cannot be explained by channel/lead numbers or amplitude effects alone.

Conclusion: The techniques developed are essential, because of the many extra MCG channels to analyse, and will encourage the use of MCG facilities.

Publication

Smith FE, Langley P, van Leeuwen P, Hailer B, Trahms L, Steinhoff U, Bourke JP, Murray A. Comparison of magnetocardiography and electrocardiography. Anadolu Kardiyol Derg. 2007 Jul;7 Suppl 1:20-2.

Contact

Dr. Fiona E. Smith, Freeman Unit, Newcastle upon Tyne. Tel: +44 (0)191-223-1370.

Top of page

W3C XHTML 1.0 logo
W3C CSS logo
Triple-A conformance icon, W3C-WAI Web Content Accessibility Guidelines 1.0
Internet Content Rating Association logo

Regional Medical Physics Department, Newcastle General Hospital, Newcastle upon Tyne, NE4 6BE, UK.
Tel: +44 (0)191-233-6161 Fax: +44 (0)191-226-0970 E-mail: webmaster@rmpd.org.uk Web: http://www.rmpd.org.uk
By using this site you agree to our Terms and Conditions of Use. Please read our legal page.
Regional Medical Physics Department, a Clinical Directorate of the Newcastle upon Tyne Hospitals NHS Foundation Trust.
Copyright ©1999-2008 All Rights Reserved. Revised 28 April, 2008.

This Web Site

Accessibility | Feedback | Help | Links | Site Map | What's New |

Our Organisation

FAQs | Find Us | News & Events | Services | Work For Us |

Our Research

Abstracts | Innovation | Publications | What We Do |

Quick Site Search

Advanced search