Raumedic Neurovent-P ICP sensor

Background: In clinical practice, fibreoptic and piezo-electric ICP probes are often used for measuring intracranial pressure (ICP). A number of similar technologies, although performing well in bench test studies, have been shown to exhibit unacceptable zero drift, fragility or both during trials conducted under clinical conditions. Recently, a new technology has become available, the Neurovent-P (Raumedic AG+CO, Raumedic, Germany). As a pre-requisite for a clinical trial, we have conducted and report on bench test studies to confirm the manufacturer's long term zero-drift performance for this technology.

Method: In a test rig static tests (recording of 20mmHg pressure) and dynamic tests, ranging from 5 to 50mmHg have been performed.

Findings: 10 ICP probes have been tested for a total of 60 days. All the catheters, after the connection with the ICU monitor displayed a static pressure of 0 ± 1mmHg and did not required pre-insertion alteration. At five days, mean zero drift was 0.6 ± 0.9mmHg. Overall, zero drift ranged from 0 to 2mmHg. At a fixed static pressure of 20mmHg, the mean recorded value was 20.6 ± 0.8mmHg, ranging from 19 to 23mmHg. A regression analysis of the relationship between the applied pressure and the recorded pressure during the dynamic tests of the 10 catheters yielded a correlation coefficient R² of 0.997. Applying the Altman and Bland method to assess the bias and confidence limits for the Raumedic catheter responses during the dynamic tests against the applied gold-standard hydrostatic column pressures, the average bias of -0.66 ± 0.85mmHg, with 95% CLs of -2mmHg and 1mmHg.

Conclusions: Mean zero drift, after five days, was very small and long-term continuous recording of a stable pressure was very precise. The response at dynamic tests, i.e. the changes of pressure in a wide range, was excellent. The average bias of the Raumedic catheter compared with the hydrostatic column is very small. After this bench test, the next and most critical step will be to conduct a trial of this promising technology under more demanding clinical environment.

Publication

Citerio G, Piper I, Cormio M, Galli D, Cazzaniga S, Enblad P, Nilsson P, Contant P, Chambers I, on behalf of the BrainIT Group. Bench test assessment of the new Raumedic Neurovent-P ICP sensor: a technical report by the BrainIT group. Acta Neurochirurgica, Volume 146, Number 11. Pages: 1221 - 1226.

Contact

Dr. Iain R. Chambers, Cleveland Unit, Middlesbrough. Tel: +44 (0)1642-854-778.

Top of page

W3C XHTML 1.0 logo
W3C CSS logo
Triple-A conformance icon, W3C-WAI Web Content Accessibility Guidelines 1.0
Internet Content Rating Association logo

Regional Medical Physics Department, Newcastle General Hospital, Newcastle upon Tyne, NE4 6BE, UK.
Tel: +44 (0)191-233-6161 Fax: +44 (0)191-226-0970 E-mail: webmaster@rmpd.org.uk Web: http://www.rmpd.org.uk
By using this site you agree to our Terms and Conditions of Use. Please read our legal page.
Regional Medical Physics Department, a Clinical Directorate of the Newcastle upon Tyne Hospitals NHS Foundation Trust.
Copyright ©1999-2008 All Rights Reserved. Revised 28 April, 2008.

This Web Site

Accessibility | Feedback | Help | Links | Site Map | What's New |

Our Organisation

FAQs | Find Us | News & Events | Services | Work For Us |

Our Research

Abstracts | Innovation | Publications | What We Do |

Quick Site Search

Advanced search