In the absence of reliable supplies of X-ray film, direct fluoroscopy is still extensively used in Albania, with chest radiology a particularly common application. This paper aims to quantify both patient skin dose and the risk-related quantity effective dose for direct fluoroscopy units based in seven different Albanian X-ray departments.

A standard Quality Assurance (QA) protocol was used to assess tube potential accuracy, half value layer and X-ray tube output of these units. Three groups of X-ray beam parameters were defined from the QA results. covering the range of chest posteroanterior (PA) fluoroscopy technique factors seen during the study. Organ-equivalent doses were then measured for a nominal PA chest fluoroscopy examination using a Rando anthropomorphic phantom loaded with lithium fluoride thermoluminescent dosimeter chips.

Normalised organ doses are listed for the three groups of beam conditions simulated. Using these factors, effective dose for the seven systems surveyed was found to be between 0.06 and 0.42 mSv for a 20 s PA chest fluoroscopy examination. Mean effective dose for this group of systems was 0.22 mSv which is a factor of 13 greater than mean effective dose for film/screen PA chest radiography in the UK, whereas entrance surface dose was a factor of 50 greater than the current EU reference level.

Publication

Marshall NW, Shehu G, Marsh D, Faulkner K, Malone J, Dewhurst K. Effective dose in Albanian direct chest fluoroscope. Eur. Radiol (2001)11:705-710.

Contact

Dr. Nick Marshall, NGH Unit, Newcastle upon Tyne. Tel: +44 (0)191-233-6161.

Top of page

W3C XHTML 1.0 logo
W3C CSS logo
Triple-A conformance icon, W3C-WAI Web Content Accessibility Guidelines 1.0
Internet Content Rating Association logo

Regional Medical Physics Department, Newcastle General Hospital, Newcastle upon Tyne, NE4 6BE, UK.
Tel: +44 (0)191-233-6161 Fax: +44 (0)191-226-0970 E-mail: webmaster@rmpd.org.uk Web: http://www.rmpd.org.uk
By using this site you agree to our Terms and Conditions of Use. Please read our legal page.
Regional Medical Physics Department, a Clinical Directorate of the Newcastle upon Tyne Hospitals NHS Foundation Trust.
Copyright ©1999-2008 All Rights Reserved. Revised 28 April, 2008.

This Web Site

Accessibility | Feedback | Help | Links | Site Map | What's New |

Our Organisation

FAQs | Find Us | News & Events | Services | Work For Us |

Our Research

Abstracts | Innovation | Publications | What We Do |

Quick Site Search

Advanced search