Comparison of patient dose

Objectives: To compare the radiation doses from imaging protocols for dental implant planning either using conventional radiography only (dental panoramic radiography (DPR), cephalometry and linear cross-sectional tomography) or involving computed tomography (CT).

Methods: Organ absorbed doses were measured using a female Rando anthropomorphic phantom loaded with lithium fluoride thermoluminescent dosemeters (TLD). Standard mandibular protocols for dental implant planning were followed using either a conventional dental radiographic unit (PM 2002 CC Planmeca, Helsinki, Finland) or CT scanner (Excel Twin Elscint, Haifa, Israel). Organ absorbed and effective doses were calculated. Effective dose was calculated using two approaches, one based on the ICRP method which excludes the salivary tissue from the remainder organs (designated Eexc) and the other with its inclusion (Einc).

Results: The greatest individual organ doses for any examination were measured in the salivary tissue. Eexc for panoramic, cephalometric and cross-sectional tomography using DPR was 0.004 mSv, 0.002 mSv and 0.002 mSv, respectively, whereas with CT it was 0.314 mSv. The value of Einc calculated using these data was between two and five times Eexc.

Conclusions: Einc greatly increases the apparent radiation burden, especially with high dose procedures. CT techniques can provide excellent images, but at the cost of increased radiation detriment. DPR with a cross-sectional tomography facility may give adequate clinical information at a greatly reduced dose.

Publication

Lecomber AR, Yoneyama Y, Lovelock DJ, Hosoi T, Adams AM. Comparison of patient dose from imaging protocols for dental implant planning using conventional radiography and computed tomography. Dentomaxillofacial Radiology (2001) 30, 255-259.

Contact

Dr. Andy Lecomber, NGH Unit, Newcastle upon Tyne. Tel: +44 (0)191-233-6161

Top of page

W3C XHTML 1.0 logo
W3C CSS logo
Triple-A conformance icon, W3C-WAI Web Content Accessibility Guidelines 1.0
Internet Content Rating Association logo

Regional Medical Physics Department, Newcastle General Hospital, Newcastle upon Tyne, NE4 6BE, UK.
Tel: +44 (0)191-233-6161 Fax: +44 (0)191-226-0970 E-mail: webmaster@rmpd.org.uk Web: http://www.rmpd.org.uk
By using this site you agree to our Terms and Conditions of Use. Please read our legal page.
Regional Medical Physics Department, a Clinical Directorate of the Newcastle upon Tyne Hospitals NHS Foundation Trust.
Copyright ©1999-2008 All Rights Reserved. Revised 28 April, 2008.

This Web Site

Accessibility | Feedback | Help | Links | Site Map | What's New |

Our Organisation

FAQs | Find Us | News & Events | Services | Work For Us |

Our Research

Abstracts | Innovation | Publications | What We Do |

Quick Site Search

Advanced search