Quality assurance for MRI

The aim of this study is to propose guidelines for quality assurance (QA) in MRI, based on a comprehensive assessment of QA parameters undertaken on a busy clinical MRI scanner over the course on one year. QA phantoms supplied by the scanner manufacturer were used together with the Eurospin MRI phantom set.

Signal-to-noise ration (SNR) and uniformity were measured daily from spin echo images acquired using a quadrature send-receive head coil and from a gradient echo sequence using the Helmholtz body coil. The voltage of the transmit radiofrequency (RF) amplifier was noted. Monthly measurements of the slice thickness, geometric distortion, slice position, image resolution and image ghosting were acquired using the head coil. In addition, SNR was measured monthly on a selection of commonly used coils. Apart from some drift of the RF amplifier voltage, all measurements were within acceptable limits and were stable over one year.

Satisfactory measurements of SNR were possible using the simple phantom supplied with the scanner. The SNR, geometric distortion and RF amplifier voltage are simple to determine and can be measured in less than 15 min by the scanner operator, using the scanner software. Weekly recording of these parameters is recommended for busy clinical MRI scanners, as this should allow deviations from acceptable limits to be identified early. Such in-house checks can be usefully compared with the less frequent estimations performed by the service engineer. Comprehensive QA routines are discussed for systems used for quantitative measurements.

Publication

Firbank MJ, Harrison RM, Williams ED, Coulthard A. Quality assurance for MRI: practical experience. British Journal of Radiology, 73 (2000) 376-383.

Contact

Dr. Michael J. Firbank, RVI Unit, Newcastle upon Tyne. Tel: +44 (0)191-233-6161.

Top of page

W3C XHTML 1.0 logo
W3C CSS logo
Triple-A conformance icon, W3C-WAI Web Content Accessibility Guidelines 1.0
Internet Content Rating Association logo

Regional Medical Physics Department, Freeman Hospital, Newcastle upon Tyne, NE7 7DN, UK.
Tel: +44 (0)191-233-6161 E-mail: webmaster@rmpd.org.uk Web: http://www.rmpd.org.uk
By using this site you agree to our Terms and Conditions of Use. Please read our legal page.
Regional Medical Physics Department, a Clinical Directorate of the Newcastle upon Tyne Hospitals NHS Foundation Trust.
Copyright ©1999-2008 All Rights Reserved. Revised 7 November, 2008.

This Web Site

Accessibility | Feedback | Help | Links | Site Map | What's New |

Our Organisation

FAQs | Find Us | News & Events | Services | Work For Us |

Our Research

Abstracts | Innovation | Publications | What We Do |

Quick Site Search

Advanced search