Varying MLC end projection size

Siemens linacs use multileaf collimators (MLCs) that move in an arc such that the flat faces of the leaf ends are always in the same plane as the radiation focus. An effect of this is that the magnification of the leaf end projection the isocentric plane normal to the collimator rotation axis varies with the drive position of the leaves. This effect is quantified theoretically and empirically verified. A method is introduced for assessing the importance of the effect for a particular MLC pattern. The significance of the effect is discussed.

Publication

Buckle AH. Varying MLC end projection size: an effect of the double-focused Siemens multileaf collimator. Phys. Med. Biol. 51 (2006) N147-N155.

Contact

Andrew Buckle, NGH Unit, Newcastle upon Tyne. Tel: +44 (0)191-233-6161.

Top of page

W3C XHTML 1.0 logo
W3C CSS logo
Triple-A conformance icon, W3C-WAI Web Content Accessibility Guidelines 1.0
Internet Content Rating Association logo

Regional Medical Physics Department, Freeman Hospital, Newcastle upon Tyne, NE7 7DN, UK.
Tel: +44 (0)191-233-6161 E-mail: webmaster@rmpd.org.uk Web: http://www.rmpd.org.uk
By using this site you agree to our Terms and Conditions of Use. Please read our legal page.
Regional Medical Physics Department, a Clinical Directorate of the Newcastle upon Tyne Hospitals NHS Foundation Trust.
Copyright ©1999-2008 All Rights Reserved. Revised 7 November, 2008.

This Web Site

Accessibility | Feedback | Help | Links | Site Map | What's New |

Our Organisation

FAQs | Find Us | News & Events | Services | Work For Us |

Our Research

Abstracts | Innovation | Publications | What We Do |

Quick Site Search

Advanced search